光学 Waves & Optics · Laboratory experiment

Verification of the Beer–Lambert law (UV–visible spectrophotometry)

Aim: to prepare a dilution series from a stock solution, record its absorption spectrum to find λmax, measure absorbance against concentration, verify A = εcl, determine the molar extinction coefficient ε, find an unknown concentration, and explore the instrumental and chemical deviations from the law.

Spectrophotometer

Prepared solutions

Absorption spectrum of the cuvette

Observation table

Calibration graph: A against c

Results

Theory

A slab of thickness dx removes the fraction \(dI/I = -\sigma n\,dx\) of the light passing through it. Integrating gives the Beer–Lambert law

$$ T = \frac{I}{I_0} = 10^{-\varepsilon c l},\qquad A = -\log_{10}T = \varepsilon\,c\,l ,$$

with ε the molar extinction coefficient (L mol⁻¹ cm⁻¹), c in mol/L and l in cm. A plot of A against c at λmax is a straight line through the origin with slope εl.

Why measure at λmax?

There the sensitivity is highest and dA/dλ = 0, so small wavelength errors and the finite spectral bandwidth matter least. Try measuring on the steep flank of a band with a wide slit.

Deviations from the law (all simulated)

  • Stray light s: \(A_{meas} = -\log_{10}\frac{T+s}{1+s}\). A saturates near \(-\log_{10}s\), so the calibration curves over at high A.
  • Polychromatic light: with a bandwidth comparable to the band width, an average of \(10^{-A(\lambda)}\) is measured instead of A(λ), giving negative curvature.
  • Chemical: methylene blue dimerises (2M ⇌ D, K ≈ 4000 M⁻¹). The dimer absorbs at 605 nm rather than 664 nm, so A(664) rises more slowly than c.
  • Photometric noise is worst at very high and very low A. The relative error in c is smallest near A ≈ 0.43 (T = 37%).

Procedure

  1. Choose a solution and scan its spectrum to locate λmax (KMnO₄ 525 nm, CuSO₄ 810 nm, K₂Cr₂O₇ 350 nm, methylene blue 664 nm).
  2. Set λ = λmax, insert a cuvette of pure solvent and press Blank. Without a blank, the cuvette and solvent absorption (about 0.04) offsets every reading.
  3. Prepare a series (1, 2, … 10 mL of stock diluted to 10 mL) and record A for each.
  4. Plot A against c and fit within the linear range. Then ε = slope / l.
  5. Make an unknown, measure it, and read its concentration from the calibration line.

Reference values

  • KMnO₄: ε525 ≈ 2.45 × 10³ L mol⁻¹ cm⁻¹ (vibronic peaks at 507, 525, 546, 566 nm)
  • K₂Cr₂O₇ (dilute HClO₄): ε350 ≈ 3.15 × 10³, ε257 ≈ 4.8 × 10³
  • CuSO₄(aq): ε810 ≈ 12 L mol⁻¹ cm⁻¹ (a d–d band, very weak)
  • Methylene blue: ε664 ≈ 7.4 × 10⁴ (monomer)

Viva questions

Why is absorbance additive but transmittance is not?

Transmittances of successive layers multiply, so their logarithms add. A is proportional to the number of absorbers in the beam.

Why use a blank?

It cancels reflection at the cuvette windows and absorption by the solvent, so only the solute contributes to A.

What is the colour of KMnO₄ and why?

It absorbs green light (500–560 nm) and transmits red and blue, so it looks purple. The cuvette drawing above uses the transmitted spectrum to set its colour.